Figure 2-9.  CC PARD Test Setup
Initialization Procedure
Follow the procedure if either the GPIB assembly has been replaced, or the EEPROM (U70) has been replaced:
1.  Install the GPIB assembly in the unit.
2.  Turn the power on and depending on your unit's model number, send string:
 
"EEINIT 6030"
 
"EEINIT 6031''
 
"EEINlT 6032"
 
          or
 
''EEINIT 6035"
3.  Turn the power off, wait 5 seconds, then turn the power back on.
4.  If the GPIB assembly has been replaced, calibrate the unit.