AudioQuest M-27 Manuale Utente

Pagina di 25
L-AVG                The average or equivalent level for the HTL, LTL, or
 
LEQ                     3 dB for the event. 
 
DOSE                  The percent Dose for the HTL, LTL, or 3 dB for the
 
event. 
 
TWA                   The Time Weighted Average for HTL or LTL or the
 
SEL                      Sound Exposure Level for the event. 
 
PEAK                 The Peak Level for the event.
 
LEVEL
 
 
MAX                   The Maximum Level for the event.
 
LEVEL
 
 
RUN                    The Run Time of the event.
 
TIME
 
 
Setting the Clock 
If the clock is not set, it functions as an elapsed time clock from the time the unit is reset.
 
 
The clock can be either a 12- or 24-hour clock depending on an internal switch.
 
 
When the clock is set to the current time, it will record the actual time for Events and Histograms.
 
CODE                  Press until the display includes a colon. The display
 
HL3                     is in hours and minutes. To set clock, press 0 and num-
 
 
 
bers for the current time. The time will move in from the right. The clock is not set until  
PAUSE/RESET is pressed. 
 
PAUSE               Resets the current time to the number in the display.
 
RESET                 The display then shows CLo. 
 
Recording the Calibration Level 
The procedure does not calibrate the unit. It stores the CAL level to assure the unit was calibrated.
 
CODE                 Press the key until CAL and the current calibration
 
HL3                    level is displayed. To recalibrate the unit, press SOUND
 
LEVEL with the microphone inserted in a calibrator.
 
SOUND              The display will show SPL (the calibrator level). Wait
 
LEVEL               until the level is steady and press PAUSE/RESET.
 
If the unit is slightly out of calibration, gently adjust 
 
the CAL screw. If the unit is more than a few decibels out of calibration, it should be 
rechecked. 
 
PAUSE               Resets the calibration level. Hold key down until the
 
RESET               display shows CAL. 
 
Using a Printer 
The printout (Figure 2) can include a heading, a summary of the data, event data, 1 to 3 
histograms, 1 or 2 percent time statistical distributions, and 1 or 2 percent dose statistical distributions. The 
histograms can be of 1, 3, 5, or 10-minute duration. The histograms and statistical distributions can be text, 
graphical or both. The printout can be changed and edited before or during printing to print only desired 
data. Data is not lost during a printout. 
 
CODE                 To print only HTL, LTL, or 3 dB histograms and statis-
 
HL3                      tical distributions, hold key down until the one desired
 
is displayed. To print all 3, hold key down until all 3
 
are displayed. 
 
PRINT                Prints heading, summary, events, and desired histo-
 
grams and distributions. 
 
HIST 1                Prints 1-minute Histograms or 5 minutes per line. If
 
HTL, LTL, and 3 dB are displayed all 3 will be printed. 
 
HIST 3                Prints 3-minute Histograms or 15 minutes per line. 
 
HIST 5                Prints 5-minute Histograms or 30 minutes per line. 
 
HIST 10              Prints 10-minute Histograms or 1 hour per line. 
 
If HIST 1,3,5, or 10 is pressed during a printout, the next line will print with the new time base for 
a clearer report. 
 
STAT                   Prints a % Time and a % Dose statistical distribution. 
 
PAUSE                Pauses the printer to edit printout or adjust printer.
 
RESET                The display shows PSE. To continue printing, press
 
PAUSE again. Other print keys will alter the printout.
 
To conserve power do not leave PSE on the display.
 
ON/OFF will also exit print mode. 
 
ON                       Stops the printout. If the printer buffer contains data
 
OFF                      the printer may run for several seconds after O-5 is
 
displayed; however, the printout has stopped.