A&D GH-120/GH-200/GH-300/GH-202/GH-252 Manuale Utente

Pagina di 86
 
 56 
 
12.3. 
Data Memory for Calibration and Calibration Test 
 Calibration 
data 
(when 
and how it is performed) and calibration test 
data can be stored in memory. 
  All the data in memory is available to be output at one time to a 
printer or personal computer.   
  Up to 50 data sets of the latest calibration or calibration test can be 
stored. When the memory capacity has been reached, "
fUl
illuminates in the upper left of the display as shown below.   
 
 
Storing the calibration and calibration test data 
Note If "
Clr
" appears blinking in the upper left of the display, delete the data in memory. 
 
Store the calibration and calibration test data as follows: 
1  Set the "Data memory (
data
)" parameter to "
2
". 
2  Set the "GLP output (
info
)" parameter to "
1
" or "
2
". 
3  With the settings above, each time calibration or calibration test is performed, 
the data is stored automatically. 
 
 
Transmitting the memory data 
Note 
  Confirm that the "Serial interface (
5if
)" parameters are set properly.   
Refer to "10. Function Table" and "15.2. Connection To Peripheral Equipment".   
  Confirm that the "Data memory (
data
)" parameter is set to "
2
". 
1  Press and hold the 
 PRINT 
 key until 
 out 
  is displayed, then 
release the key. 
2 Press 
the 
 PRINT 
 key to display 
 out "n"o 
 with "no" blinking. 
3 Press 
the 
 RE-ZERO 
  key to display 
 out "g"o 
 with "go" blinking. 
4 Press 
the 
 PRINT 
  key to transmit all memory data using the RS-232C interface. 
5  The balance displays 
 Clear 
  when all memory data is output.   
Press the 
 CAL 
  key to return to the weighing mode. 
 
 
Deleting data stored in memory 
1  Press and hold the 
 PRINT 
 key until 
 out 
  is displayed, then release the key. 
2 Press 
the 
 SAMPLE 
  key to display 
 Clear 
3 Press 
the 
 PRINT 
 key to display 
 Clr "n"o 
 with "no" blinking. 
4 Press 
the 
 RE-ZERO 
  key to display 
 Clr "g"o 
 with "go" blinking. 
5 Press 
the 
 PRINT 
  key to delete all data. 
6  The balance displays 
 out 
 and returns to the weighing mode 
when all the data has been deleted. 