SECTION 5 —OPERATING INSTRUCTIONS
24
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 24.
1. Turn the flowrate knob to the setting prescribed by your 
physician or therapist.
 WARNING
DO NOT change the L/min. setting on the flowmeter 
unless a change has been prescribed by your physician or 
therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate 
line on the flowmeter. Next, turn the flow knob until the ball rises to 
the line. Now, center the ball on the L/min. line prescribed. 
2. If the flowrate on the flowmeter ever falls below 0.5 L/min. 
for Platinum 5 (1 L/min. for Platinum 10) for more than 
about one minute, the LOW FLOW alarm will be 
triggered. This is a rapid beeping of the audible alarm. 
Check your tubing or accessories for blocked or kinked 
tubing or a defective humidifier bottle. After rated flow is 
restored, more than 0.5 L/min. for Platinum 5 (1 L/min. for 
Platinum 10), the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand 
and the HomeFill II compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball