Invacare 5 Manuale Utente
SECTION 5 —OPERATING INSTRUCTIONS
24
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 24.
1. Turn the flowrate knob to the setting prescribed by your
1. Turn the flowrate knob to the setting prescribed by your
physician or therapist.
WARNING
DO NOT change the L/min. setting on the flowmeter
unless a change has been prescribed by your physician or
therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate
line on the flowmeter. Next, turn the flow knob until the ball rises to
the line. Now, center the ball on the L/min. line prescribed.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min.
for Platinum 5 (1 L/min. for Platinum 10) for more than
about one minute, the LOW FLOW alarm will be
triggered. This is a rapid beeping of the audible alarm.
Check your tubing or accessories for blocked or kinked
tubing or a defective humidifier bottle. After rated flow is
restored, more than 0.5 L/min. for Platinum 5 (1 L/min. for
Platinum 10), the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand
and the HomeFill II compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball