Teledyne 2002 Manuale Utente

Pagina di 38
149-072011 Model 2002 Vacuum Gauge
 
Page 25 of 38 
 
When a lower case character is present in an example it represents an option. 
Character 
 
Description  
 
 
 
 
 
 
Valid Inputs: 
RS-485 Address (hexadecimal 0-9, A-F)........................................................01 - FF 
Most Significant Digit Of Mantissa..................................................................... 1 - 9 
d Decimal 
Digit ...................................................................................................... 0 - 9 
e Exponent ............................................................................................................. 0 - 5 
Unit Of  Pressure........................................................................................T, M, or P 
, Command 
Separator (comma) ............................................................................ N/A 
<CR> 
Command Terminator (carriage return).............................................................. N/A 
 
Interrogation Commands 
 
Command Description   
 
Format   
 
 
 
Sample Response 
 
Transmit Averaged Pressure 
P<CR> ................................................ Pa: 1.23456e+0 Torr<CR> 
Transmit Pirani Pressure 
R<CR>..................................................Pr: 1.98765e-3 Torr<CR> 
Transmit Piezo Pressure 
Z<CR>................................................ Pz: 7.65432e+2 Torr<CR> 
Transmit RS-485 Address 
A<CR>.............................................. Multidrop Address: 01<CR> 
Transmit Decimation Ratio 
D<CR> ............................................. Decimation Ratio: 255<CR> 
Transmit Selected Gas # 
G<CR> .....................................................................Gas#: 0<CR> 
Transmit High Set point 
H<CR> ............................................... Hi: 1.00000e+1 Torr<CR> 
Transmit Low Set point 
L<CR>................................................. Lo: 1.00000e-2 Torr<CR> 
Transmit Device Status S<CR> ........................................................................ 00044<CR> 
Transmit Turnaround Delay T<CR> ........................................................Comm 
Delay: 6<CR> 
Transmit Selected Units U<CR> ..........................................................................Torr<CR> 
Transmit Software Version# 
V<CR>.................................... Hastings Instruments - Model 2002 
 
..................................................Version 1.60 - (07-02-2002)<CR> 
Parameter Modification Commands 
Command 
Description 
 Format 
     Valid 
Range: 
Modify High Set point 
H={m.dd}E{+e}<CR> ...................................... 1.00000e-9  to  9.99999e+9 
Modify Low Set point 
L={m.dd}E{-e}<CR> ........................................ 1.00000e-9  to  9.99999e+9 
Modify Selected Gas # 
G={d}<CR> 
0 To 4  ............................ (Decimal) 
Modify Selected Units 
U={u}<CR> 
T, or M,  ............{or P, if Available}  
Modify Decimation Ratio 
D={dddd}<CR> 
63 To 7936 ..................... (Decimal) 
Modify RS-485 Address 
*{aa}A={aa}<CR> 
1 To FF ................... (Hexadecimal)  
Modify Turnaround Delay 
*{aa}T={dd}<CR> 
0 To 255 ......................... (Decimal)