Philips 862474 ユーザーズマニュアル

ページ / 182
Diagnostic Mode
Configuring the Power-up Defaults Menu  
5-5
Diagnostic Mode
The purpose of the Diagnostic Mode Menu option is to allow factory, field-service, and 
hospital biomedical technicians access to a series of test and system-related information 
screens for the purpose of verifying monitor performance, or troubleshooting problems.
To access Diagnostic Mode:
The Diagnostic Menu lists the test and system-related information screens. Select an item in 
the menu to invoke a test or information screen. The test and information screens that are in 
the diagnostic menu listed below. An explanation of each menu item is given in the sections 
that follow.
Error Codes
System Information
System A/D Values
NBP  Test
CO
2
 Test
Step
Action
1
Enter  the  Power-Up Defaults Menu (see ““Power-Up Defaults Menu” on 
page 5-3
for instructions on accessing this menu).
2
Select the Enter Diagnostic Mode? option. 
3
From the popup menu, select Yes. The Diagnostic Menu displays.
      
           
DIAGNOSTIC MENU
Error Codes
System Information
System A/D Values
NBP Test
CO2 Test
Return