Philips 862474 ユーザーズマニュアル

ページ / 182
Part 3 Using Support Functions
6-12 
Troubleshooting  
Error Code 
Categories
As a reference, the following table lists the general categories for other error codes. The error 
code categories are shown only in hexadecimal format.
64
The SpO
2
 module is sending an 
error message to the host CPU.
1. Ensure PicoSat module is 
properly connected.
2. Replace the PicoSat module.
3. If problem persists, replace main 
PCB module.
?
The  CO
2
 module is sending an 
error message to the host CPU.
1. Ensure Oridion module is 
properly connected.
2. Rplace the Oridion module.
3. If problem persists, replace main 
PCB module.
65, 66
PicoSat SpO
2
 module has detected 
an error during initialization. 
1. Replace PicoSat module.
2. If problem persists, replace main 
PCB module.
6E-71
PicoSat SpO
2
 module has detected 
an error on its serial port. 
1. Replace PicoSat module.
2. If problem persists, replace main 
PCB module.
Hex Code
Explanation
Recommended Action
Hexadecimal Code
Explanation
500xxx
internal user interface error
501xxx
remove serial port error
502xxx
date and time error
503xxx
NBP error
504xxx
front end error
505xxx
alarm error
506xxx
audio error
507xxx
recorder error
508xxx
trend error
509xxx
flash memory data error
50axxx
SpO
2
 error
50bxxx
ECG error
50cxxx
power-down task error
50dxxx
on-board diagnostic error
50exxx
power monitor error
50fxxx
temperature measurement error
510xxx
internal user interface error
511xxx
error handling error
513xxx
serial driver error