Disassembly Procedures
7-14 
Disassembly  
Removing the
SpO
2
 Module
After you have removed the Patient Monitoring I/O module from the Main PCB module, you 
can remove the SpO
2
 module. To do this:
Step
Action
1
Remove the 2 screws fastening the SpO2 module to the I/O module.
2
Remove the connector that attaches the SpO
2
 module to the I/O module.
3
Gently pull the SpO
2
 module away.
Screw #1
Screw #2