Disassembly Procedures
Disassembly   
7-17
Removing the
Power Supply
Module
Step
Action
1
Use the needle-nose pliers to unplug the three connectors.
2
Remove the two screws that connect the power supply to the rear panel.
3
Remove the screws on the side and bottom of the rear panel.
4
Cut the tie wrap that holds the battery cables together.
5
Use a 5/16 nut driver to unfasten the power supply from the rear panel.
6
Pull the power supply module away from the rear panel.
Connector #1
Connectors #2 & 3
Power Supply Board
Battery Cables
Communications Module