Philips 862474 ユーザーズマニュアル

ページ / 182
Disassembly
11-12 
Training  
Whenever the monitor has been opened, there are specific steps regarding reassembly, 
described in the disassembly instructions in Chapter 7, that must be followed. If these 
instructions are not followed, various errors may result. 
For monitors that are defective on arrival, first verify that the problems are not related or 
cannot be remedied by following the steps in the above troubleshooting table. If these are not 
successful, and a replacement monitor is necessary, contact a Philips Sales Representative to 
arrange a replacement.
Disassembly
Disassembly procedures are described in the disassembly guide in Chapter 7. Read the whole 
chapter. 
Whenever the monitor has been opened, there are specific steps regarding reassembly, 
described in the disassembly instructions in Chapter 7, that must be followed. If these 
instructions are not followed, errors may result.
If the monitor fails the required performance tests described in Chapter4, then please review 
the reassembly.
Audible alarm does not 
sound
Speaker malfunction or speaker 
has a loose connector.
Main PCB module malfunction.
1. Check speaker connector.
2. Replace speaker according to the 
instructions in 
.
Replace the Main PCB module according 
to the instructions in 
.
No sound from monitor
Tone is switched off or low.
1. Turn the volume up by pressing the 
Volume button.
2. Power-cycle monitor. Tone should 
sound when monitor boots up. 
3. Check cable connections to the 
speaker.
No tone alarms
Speaker will not work
Audible alarms are not switched 
on.
Cable disconnected.
Defective speaker.
Make sure audible alarms are switched on.
Reconnect cable.
Replace speaker.
The wheel is rotated, no 
highlight appears on the 
display screen, and/or the 
monitor does not respond to 
wheel presses.
Main PCB module malfunction.
Replace the Main PCB module according 
to the instructions in 
.