Site Preparation Guidelines
Site Preparations  
2-3
Site Preparation Guidelines
Check that the environment is reasonably free from dust, vibration, corrosive or 
explosive gases and is within specified temperature and humidity ranges. See 
Chapter 12, “Specifications”
Position the monitor so that the display is clearly visible and the controls are easy to 
reach.
Warning, Cautions and Safety Precautions
Patient 
Safety
To better secure patient safety, become familiar with the details in 
Chapter 12, “Specifications.”  
Patient 
Leakage 
Current
The patient leakage current is <= 50µA @ 250 V. The monitor has patient input connectors 
(Type CF) that are protected against the effects of defibrillation and electrosurgery.
You will find the following heart symbol on the side of the monitor where the patient 
monitoring input connectors are located. The symbol indicates that the monitor is Type CF 
and is designed to have special protection against electric shocks (particularly regarding 
allowable leakage currents, having a F-Type applied part, according to the standards IEC 
60601-1/EN60601-1/CSAC22.2 601.1/UL 2601-1), and is defibrillator proof.