Warnings, Cautions, and Safety Precautions Relating to Wall Mount Installation
2-8 
Site Preparations  
Warnings, Cautions, and Safety Precautions Relating to Wall Mount 
Installation
Make sure that you have read all applicable instructions before attempting to install the 
wall mount.
Wall mounts that are intended to support monitors must be capable of supporting four 
times the weight of the monitor when properly installed. 
If wall mounts are to be installed on plaster board walls, we recommend that the mounts 
be installed using steel or plywood plates sufficient to distribute the load over a large 
section of the wall. Lead expansion bolts are neither adequate nor safe for this 
purpose
.
It is the responsibility of the hospital, institution, or its designated representatives to 
determine that the wall is adequate to safely mount monitoring instrumentation. This 
includes the selection of and proper installation of the appropriate fasteners and mounts. 
In a new construction, or in a remodelled construction where the wall can be completely 
or partially removed, a 16-gauge steel stud should be mounted for the purpose of 
attaching the wall channel. This design safety factor is based on four times (4x) the 
maximum rated load.
GCX Wall Channel: Always reposition the adjustable channel safety stop under the 
adapter plate when the height of the monitor is changed.
The channel safety stop prevents the mounting device from inadvertently falling or 
being pulled out of the wall channel. Do not remove this stop while the mount is 
supporting a monitor
.
Do not mount any portion of the monitor over a patient’s bed.
Do not exceed the maximum rated load specified for each wall mount.
Ensure that no electrical wiring, utilities, or piping interface with the selected wall 
mounting location.
Do not install wall channel onto solid brick or brick veneer walls. Do not attempt to 
install wall mounts onto crumbly wall material.
Check the mounting hardware holding the wall channel or bracket to the wall every 12 
months. Tighten if necessary.