Philips 862474 ユーザーズマニュアル

ページ / 182
Testing Safety Information
4-2 
Testing the Monitor  
Testing Safety Information
Warning
Warning
Perform all leakage tests any time the unit is opened.
AC mains voltage is present on the applied part terminals during this test. Exercise 
caution to avoid electrical shock hazard.
Note
The monitor must be placed in Diagnostic Mode, with the NBP test screen active for each of 
the NBP tests.
If the test fails, ensure the integrity of the cuff and tubing, then test again. If the test fails 
again, verify the integrity of all the pneumatic tubing inside the monitor.
Before beginning the Battery Performance test, ensure that the monitor is disconnected from 
the AC power source.