Functionality Assurance Tests
4-12 
Testing the Monitor  
Alarm Test
Tools Needed:
SpO
2
 adapter cable (M1943A)
Clinical Dynamics Corp - SmartSat simulator (with Nellcor simulator cable)
Volume 
Control Test
Tools Needed:
SpO
2
 adapter cable (M1943A)
Clinical Dynamics Corp - SmartSat simulator (with Nellcor simulator cable)
Step
Action
1
Connect SpO
2
 simulator cable to the SpO
2
 adapter cable. Connect the cable to 
the SpO
2
 patient monitoring input connector.
2
Set the simulator as follows: 
3
Press  the  On/Standby button to turn the monitor On.
4
After the normal power-up sequence, verify that the SpO
2
% display initially 
indicates zero, or is blank.
5
Verify that the following monitor reaction occurs:
a. After approximately 45 seconds, the monitor displays saturation and heart rate 
as specified by the tester. Verify that the values are within the following 
tolerances:
– Oxygen Saturation Range 79% to 83%
– Heart Rate Range 33 to 39 bpm
b. The audible alarm sounds and both the SpO
2
% and Heart Rate (HR) displays 
flash, indicating both parameters have violated the default alarm limits.
c. The HR tone is heard. For this test, the HR tone source must be set to SpO
2
 
from the Heart Rate Menu.
6
Press  the  Alarm Silence button to temporarily silence the audible alarm.
7
Verify the following:
a. The audible alarm remains silenced. 
b. The crossed-out bell icon displays in each numeric frame on the screen.
c. The SpO
2
% and HR displays continue flashing.
d. The HR tone remains audible.
e. The audible alarm returns in approximately 120 seconds.
Item
Setting
Oximeter
Nellcor
SpO
2
%
81
BPM
36
Pulse Mod
0.50%