Omega Engineering DOCN600 ユーザーズマニュアル
Page
2
2.0 SPECIFICATIONS
DISPLAY:
3 1/2 digit LED, 1/2" high digits
3 1/2 digit LED, 1/2" high digits
MEASURING RANGES:
D.O.: 0-20.00 ppm and 0-100% saturation, switch
selectable
Temperature: 0-40°C (32 to 104°F)
D.O.: 0-20.00 ppm and 0-100% saturation, switch
selectable
Temperature: 0-40°C (32 to 104°F)
POWER REQUIREMENTS:
98-132 Vac, 50/60 Hz (less than 10 VA)
98-132 Vac, 50/60 Hz (less than 10 VA)
Optional: 196-264 Vac, 50/60 Hz
AMBIENT CONDITIONS:
-30 to 50°C (-22 to 122°F)
0 to 90% R.H. non-condensing
-30 to 50°C (-22 to 122°F)
0 to 90% R.H. non-condensing
ANALOG OUTPUTS:
Non-isolated 0-1 mA, 100 ohms maximum load
Non-isolated 0-5 Vdc, 1000 ohms minimum load
Isolated 4-20 mA, 800 ohms maximum load
Non-isolated 0-1 mA, 100 ohms maximum load
Non-isolated 0-5 Vdc, 1000 ohms minimum load
Isolated 4-20 mA, 800 ohms maximum load
The isolated output is isolated from the input, ground
line power and all other outputs
Range Expand: The 4-20 mA analog output can be
made to represent any segment of the measuring
scale. Minimum segment is 10% of full scale.
Output Hold: The analog outputs are automatically
placed on hold during calibration or other setup
operations.
Temperature Output: The 0-5 Vdc output can be
programmed to follow either the process temperature
or dissolved oxygen.
line power and all other outputs
Range Expand: The 4-20 mA analog output can be
made to represent any segment of the measuring
scale. Minimum segment is 10% of full scale.
Output Hold: The analog outputs are automatically
placed on hold during calibration or other setup
operations.
Temperature Output: The 0-5 Vdc output can be
programmed to follow either the process temperature
or dissolved oxygen.
TEMPERATURE COMPENSATION:
Automatic 0 to 100°C (32 to 212°F)
ALTITUDE COMPENSATION:
Automatic 0 to 40°C (32 to 104°F)
DIAGNOSTICS:
When system error is indicated, call STATUS to obtain
condition code. On-board simulated input to aid in
troubleshooting to obtain condition code.
condition code. On-board simulated input to aid in
troubleshooting to obtain condition code.
MEMBRANE ALARM:
A relay activates if membrane is perforated.
TEST:
Display value and analog outputs can be set manually
Display value and analog outputs can be set manually
to any value for testing and diagnostic purposes. This
feature allows the outputs to be tested independently
of process value.
feature allows the outputs to be tested independently
of process value.
SAFETY & SECURITY:
Non-volatile memory (EEPROM)
Passcode protected if selected
Watch-dog timer monitors microprocessor
Instrument automatically returns to on-line operation if
accidentally left in menu mode
(This feature may be held disabled if desired)
Passcode protected if selected
Watch-dog timer monitors microprocessor
Instrument automatically returns to on-line operation if
accidentally left in menu mode
(This feature may be held disabled if desired)
ACCURACY: % Saturation:
± 1%
ppm:
±
0.1
Temp:
±
0.2°C
STABILITY: 0.1% of span per 24 hrs., non-cumulative
RESOLUTION: % Saturation:
RESOLUTION: % Saturation:
0.1
ppm:
0.01
Temp:
0.1°C
REPEATABILITY: 0.1% of span or better
RESPONSE TIME: 20 seconds to 90% of value upon
step change at 20°C
ENCLOSURE: NEMA 4X fiberglass reinforced polyester
enclosure with four 1/2” conduit holes and mounting
feet for surface mount
feet for surface mount
MOUNTING CONFIGURATIONS:
Standard is surface mount
Optional panel mount hardware. Part No.
Standard is surface mount
Optional panel mount hardware. Part No.
DOCN600-PM
Optional pipe mount hardware. Part No.
DOCN600-PIPE
NET WEIGHT: 3 1/2 lbs. (1.6 kg)