8
Starting therapy using oxygen
1
Fit the oxygen connector to the oxygen inlet of the device.
2
Attach the other end of the oxygen supply tubing to the oxygen supply.
3
Press
to start treatment.
4
Turn on oxygen.
Stopping therapy using oxygen
1
Turn off oxygen.
2
Press
to stop treatment.
Using the FiO
2
monitoring sensor
Use of the FiO
2
monitoring sensor may be recommended by your clinician.
CAUTION
Do not use the FiO
2
monitoring sensor with the H4i humidifier.
Preparing to use a new sensor
1
Before use, leave the FiO
2
monitoring sensor open to the air for 15 minutes.
2
Attach a new FiO
2
monitoring sensor (as shown below).
3
Perform the sensor calibration (see “Setup menu” on page 16).
Note: The FiO
Note: The FiO
2
monitoring sensor must be replaced every 12 months.