94  Stratos LV/LV-T Technical Manual   
C
AUTION
 
Overdrive Pacing Mode - When programming the overdrive 
pacing mode, check whether the selected program can cause 
PMT, and whether atrial over drive pacing would result.  
Corresponding to the measured retrograde conduction time, 
the PMT protection interval must be programmed to a correct 
value. 
2.13  AES Detection and Pacing 
Post atrial extrasystoles (AES) pacing is intended to prevent the 
occurrence of pro-arrhythmic long-short-long sequences due to 
an atrial extrasystole, which can lead to atrial tachycardias.  This 
is achieved through a shortened post AES basic interval, which 
is incrementally lengthened again during subsequent cycles until 
the basic interval is attained.   
The algorithm "Post AES Stimulation" is comprised of two parts. 
First, an AES must be recognized (AES detection), and 
secondly, the respective stimulation, Post AES Stimulation, must 
then occur. 
2.13.1 AES Detection 
To detect atrial extrasystoles (AES), an "AES (timing) window" is 
defined, the length of which is calculated from the averaging of 
the rates of the four paced or intrinsic cardiac events prior to an 
AES. The AES window is shortened by a programmable 
percentage (5... (5) ...50%) from the averaged interval (nominal 
setting is 25%).  If an intrinsic atrial event falls within this AES 
window and the coupling interval <750 ms, the cardiac event is 
classified as an AES.