VXi VT1529A/B ユーザーズマニュアル

ページ / 529
280 VT1422A Command Reference  
DIAGnostic
DIAGnostic:CUSTom:MXB
DIAGnostic:CUSTom:MXB <slope>,<offset>,(@<ch_list>) 
 sends the <slope> 
and <offset> parameters that allow the driver to calculate and download a custom 
linear Engineering Unit Conversion table to the VT1422A. Use the 
“[SENSe:]FUNCtion:CUSTom” on page 340 to link this custom EU conversion 
with channels in <ch_list>.
Parameters
Comments
<slope> specifies the linear function’s "slope":
The <offset> parameter specifies the conversion offset at zero input volts. This 
parameter is also commonly known as the "Y-intercept."
<ch_list> specifies which channels may use this custom linear function.
Related Commands: [SENSe:]FUNCtion:CUSTom (<ch_list>)
*RST Condition: All custom EU tables erased
Send with VXIplug&play Function: hpe1422_cmd(...)
DIAGnostic:CUSTom:MXB
DIAGnostic:CUSTom:MXB <slope>,<offset>,(@<ch_list>) 
 sends the <slope> 
and <offset> parameters that allow the driver to calculate and download a custom 
linear Engineering Unit Conversion table to the VT1422A. Use the 
“[SENSe:]FUNCtion:CUSTom” on page 340 to link this custom EU conversion 
with channels in <ch_list>.
Parameters
Comments
<slope> specifies the linear function’s "slope":
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
slope
numeric (float32)
limit of float32
none
offset
numeric (float32)
limit of float32
none
ch_list
channel list (string)
100 - 163
none
f
outp1
f
outp0
(
V
in1
V
in0
(
)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
slope
numeric (float32)
limit of float32
none
offset
numeric (float32)
limit of float32
none
ch_list
channel list (string)
100 - 163
none
f
outp1
f
outp0
(
V
in1
V
in0
(
)