Anritsu MG3641A ユーザーズマニュアル

ページ / 212
SECTION 4   OPERATING INSTRUCTIONS
4-23
(2) Select the item you need to change (ON/OFF or
impedance) in the "PM" line by the "
↓" [F2], and "→"
[F3] keys (the characters are highlighted).
(3) Press the "Sel" [F1] key to change the setting.
• Modulation function:  On/Off
• Impedance:  50 
Ω/600 Ω
If the pulse modulation and AM are activated simultaneously, the AM modulation will not be applied normally.
Feeding a voltage higher than +8 V or lower than 
−3 V to the "Pulse Mod Input" connector could result in
breakdown of the unit.  Be careful to prevent a voltage outside the limits from entering the unit.
CAUTION 
Modulation
[AM] 
[FM1] 
[FM2] 
[PM]
Off  Int1 
Off  Int1 
Off  Int1 
    50 
Ω
Sel
Rtn

kHz 
kHz
0.0 
0.000 
0.000
3
2
Src
Off