A&D HW-200KGL ユーザーズマニュアル

ページ / 71
HW-G Instruction Manual Page 52
HW-G OZIM v1a
16.3 Command Mode
The command mode is the function which can perform “output data”, “controlling the scale”
and “setting parameters” by a command transmitted from a computer.
Caution
Allow at least 500 milliseconds between commands.
16.3.1 Command List
The following explanation uses “format 1 ( 
 f15 0 
 )”
Data output
The current weighing data is output.
Template
Q
Command
Reply
Selection of mode and unit
Selects the mode and unit. This is the same as the 
 MODE 
 
switch.
Template
U
Command
Response
The scale changes mode and unit.
Zero
The current mass value is set to the zero point.
This is same as the 
 ZERO 
 switch.
Template
Z
Command
Response
The mass value becomes zero and zero point mark is displayed.
Tare
The current mass value is set to zero after placing a tare (container, receptacle,
case, etc) and the net is displayed. This is the same as the 
 TARE 
 
switch.
Template
T
Command
Response
The current mass value becomes zero and net mark is displayed.
Cancel of tare
The display value becomes the gross and the net mark is turned off. (The tare
value becomes zero.)
Template
CT
Command
Response
Gross is displayed and the net mark is turned off.