A&D HC-15Ki ユーザーズマニュアル

ページ / 60
50
Reply ACK C L
F
   
 
 
ID Number: 
Maximum 6 digit 
 
Upper Limit: 
Maximum 9 digit including sign 
 
Lower Limit: 
Maximum 9 digit including sign 
‰ Request the scale to reply with the setting of F-Function. 
Command  ? F C ,  0 5 0 1 C L
F
Request to reply with the setting of f-05-01.
Reply 
F C ,  0 5 0 1
,
0 C L
F
f-05-01 = 0
‰ Store the specified F-Function setting value 
Command  F C ,  0 5 0 3
,
+ 1 2 3 4 C L
F
Set “1234” as the Upper limit.
 
F-Function Number: 
4 digits 
 
Setting Value:   
Maximum 8 digits including sign. 
Reply ACK C L
F
   
 
Command O 
N C L
F
    Start with new settings. 
Reply ACK C L
F
   
 
Having finished the "FC" command, send the "ON" command to start the scale with 
new settings. The scale replies <ACK> (06H) and starts.   
            F-#        Setting value
F-#        Setting value