1-4. Setting Up Your Scale 
1. Place the scale on a suitable weighing surface (see “Best Conditions For Weighing” 
below) and turn the adjustable feet until the spirit level shows that the scale is level. 
2. Plug in the AC adapter. The AC input 
requirements could be l00, 120, 220, 
230 or 240 Volts (50/60Hz) depending 
on the area in the world so please 
check that the adapter is correct. Earth 
the chassis if you think static electricity 
may be a problem. 
3. Press the    ON/OFF    key to turn the power ON. 
All the display symbols are displayed. 
Then the display turns off once, and zero will be shown with the ZERO mark. 
4. Pressing the    ON/OFF    key again, and the power will switch OFF. 
‰
  Auto-power off function 
  It is possible to have the power automatically switched OFF. If zero is 
displayed for approximately 5 minutes. See “9-2. F-Functions” and set the 
F-Function
f-04-05
 at “
1
” to enable the function. 
5. Switch the power ON at least half an hour before use so that the scale can warm 
up.
Best Conditions For Weighing 
‰ The Scale must be level (check the spirit level on the scale). 
‰ Best operating temperature is between 20°C~25°C / 68°F~77°F at about 50%~60% 
Relative Humidity. There shouldn’t be large temperature fluctuations. 
‰ The weighing room should be kept clean and dry. 
‰ The weighing table must be of a solid construction. 
‰ Corners of rooms are best as they are less prone to vibrations. 
‰ Don’t install the scale near heaters or air conditioners. 
‰ Don’t install the scale in direct sunshine. 
‰ Try to ensure a stable AC power supply when using an adapter. 
‰ Keep equipment containing magnets away from the scale. 
‰ Warm up the scale more than 30 minutes before use. 
‰ Ground the scale chassis for electrostatic discharge if the weighing conditions 
warrant.
Calibration
Calibration of the HC-i is required when the scale is initially installed. Please see “8. 
CALIBRATION” for more calibration information. 