TARE WEIGHING 
 
Turn on scale as described above 
 
Place the “tare item” on the platform 
 
Press “ON/OFF” and wait until “0.00” is displayed 
 
Add the “net-weight-item” 
CALIBRATION 
Press “TARE” key until displays some digits on LCD after showing “8” one by one. It shows some 
digits on LCD. Press “TARE” key to calibrate zero point and show “0.00”. Put 50g weight (half 
capacity) on the platform. Press “TARE” and display “50.00”(half capacity) afterwards. Put away 
50g weight and put 100g weight (full capacity) on the platform. Press “TARE” and display “100.00”. 
Press “TARE” not release until switch off. Put away the weight. The calibration is completed. 
 
PLEASE CHECK AVAILABILITY OF WEIGHTS WITH THE STORE YOU BOUGHT THE SCALE 
FROM.