Clarity INT9 ユーザーズマニュアル

ページ / 35
Clarity - INT9 A/D converter 
 Troubleshooting 
21 
4.2  Problems with INT9 driver 
•  “Board malfunction” 
If the error message "Board malfunction
appears, the card is probably incorrectly 
inserted or damaged. 
Solution: 
Check whether the INT9 A/D card is correctly 
set in the ISA slot. The card should be tucked 
fully and should not move in the slot. If this 
does not help, contact technical support to 
replace the damaged card. 
 
Most problems with INT9 driver are generally 
indicated by following symptoms: 
•  INT9 cannot be added to 
Setup Control 
Modules
 in the 
System Configuration
 dialog, 
showing the "Cannot create detector" error 
message. 
•  Double-clicking on the device will not invoke 
the 
INT9 A/D Card Setup
 dialog. 
•  The 
Data Acquisition
 icon in the 
Instrument
 
window is not active (this problem may also 
have another cause. See section 4.4 on pg. 29). 
•  When starting Clarity  or trying to add more 
INT9 cards to configuration then is present in 
the PC, the error message “Cannot find first 
board” appears. Alternatively, “Cannot find 
second board” error message may appear. 
•  While trying to add the INT9 card in 
System 
Configuration
 dialog or while starting Clarity
"Cannot find driver file ..." error message with 
path to the INT9 drivers in the system followed 
by "Error Occurred During Setup" appeared. 
Solution: 
When one of these problems occur, check the 
INT9 driver (see next chapter) and then 
procced according to the suggested solutions.