Clarity INT9 ユーザーズマニュアル

ページ / 35
Clarity - INT9 A/D converter 
 Installation 
2.3.2 
Connection of starting cables: 
Starting input reacts to a change of the TTL 
logical level (5V) or to a connection by any 
contact (button, contact of relay). 
The input implicitly reacts to a change from 
HIGH
 to 
LOW
 (or closing of a contact). The input 
function may be altered by changing the 
Down
 
option to 
Up
 in the 
External Start/Stop
 section 
of the 
Method Setup - Measurement
 dialog (accessible 
from the 
Instrument
 window using the - 
Method - 
Measurement 
command). 
Note: 
For schemes and hints describing the typical 
autosampler wirings see the Device setup and 
wiring chapter of the Getting Started Manual. 
2.4 Clarity 
configuration 
 
1.  Start the Clarity station by clicking on the 
 
icon on the desktop. 
2.  Invoke the 
System Configuration
 dialog using the 
System – Configuration
 command. 
3.  Press the 
Add
   button (see Fig. 4) to invoke 
the 
Available Control Modules
 dialog.  
4.  Select the DataApex INT9 A/D Card and press 
the 
Add
   button. 
5.  The 
INT9 A/D Card Setup
 dialog will appear. 
 
Fig. 3. 
INT9 A/D Card Setup