Troubleshooting & Service 
Teledyne API  - T100 UV Fluorescence SO2 Analyzer  
262 
 
Sample 
Chamber
PMT 
Housing 
Mounting
Screw
Mounting 
Screw
 
Mounting 
Screw
 
 
Figure 12-7: 
Sensor Module Mounting Screws 
Follow the above steps in reverse order to reinstall the sensor module. 
12.7.2.2. CLEANING THE SAMPLE CHAMBER 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
The sample chamber should only be opened or cleaned on instructions 
from the Teledyne API Technical Support department. 
Be careful not to leave thumbprints on the interior of the sample 
chamber.  The various oils that make up fingerprints fluoresce brightly 
under UV light and will significantly affect the accuracy of the analyzer’s 
SO
2
 measurement) 