Teledyne T100 ユーザーズマニュアル
Troubleshooting & Service
Teledyne API - T100 UV Fluorescence SO2 Analyzer
262
Sample
Chamber
PMT
Housing
Mounting
Screw
Mounting
Screw
Screw
Mounting
Screw
Screw
Figure 12-7:
Sensor Module Mounting Screws
Follow the above steps in reverse order to reinstall the sensor module.
12.7.2.2. CLEANING THE SAMPLE CHAMBER
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
The sample chamber should only be opened or cleaned on instructions
from the Teledyne API Technical Support department.
from the Teledyne API Technical Support department.
Be careful not to leave thumbprints on the interior of the sample
chamber. The various oils that make up fingerprints fluoresce brightly
under UV light and will significantly affect the accuracy of the analyzer’s
SO
chamber. The various oils that make up fingerprints fluoresce brightly
under UV light and will significantly affect the accuracy of the analyzer’s
SO
2
measurement)
06807C DCN6650