Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Getting Started 
 
 
69 
Table 3-13 lists brief descriptions of the warning messages that may occur during start 
up for T100 analyzers with no options installed.  
Table 3-13: 
Possible Startup Warning Messages – T100 Analyzers w/o Options 
Message Meaning 
ANALOG CAL WARNING 
The instrument's A/D circuitry  or one of its analog outputs is not calibrated. 
BOX TEMP WARNING 
The temperature inside the T100 chassis is outside the specified limits. 
CANNOT DYN SPAN
2
 
Remote span calibration failed while the dynamic span feature was set to 
turned on. 
CANNOT DYN ZERO
3
 
Remote zero calibration failed while the dynamic zero feature was set to turned 
on. 
CONFIG INITIALIZED 
Configuration was reset to factory defaults or was erased. 
DARK CAL WARNING 
Dark offset above limit specified indicating that too much stray light is present in 
the sample chamber. 
DATA INITIALIZED 
DAS data storage was erased. 
HVPS WARNING 
High voltage power supply for the PMT is outside of specified limits. 
PMT DET WARNING 
PMT detector output is outside of operational limits. 
PMT TEMP WARNING 
PMT temperature is outside of specified limits. 
RCELL TEMP WARNING 
Sample chamber temperature is outside of specified limits. 
REAR BOARD NOT DET 
CPU unable to communicate with motherboard. 
RELAY BOARD WARN 
CPU is unable to communicate with the relay PCA. 
SAMPLE FLOW WARN 
The flow rate of the sample gas is outside the specified limits. 
SAMPLE PRESS WARN 
Sample gas pressure outside of operational parameters. 
SYSTEM RESET
1
 
The computer was rebooted. 
UV LAMP WARNING 
The UV lamp intensity measured by the reference detector reading too low or 
too high. 
1
  Clears 45 minutes after power up. 
2
 
Clears the next time successful zero calibration is performed. 
3
  Clears the next time successful span calibration is performed.  
Table 3-14 lists brief descriptions of the warning messages that may occur during start 
up for T100 analyzers with optional second gas options or alarms installed. 
06807C DCN6650