Princeton 4411-0139 ユーザーズマニュアル

ページ / 318
29
Chapter 2:
System Installation
 describes the sequence of actions required to install a PI-MAX4 system and 
prepare to gather data. Where available, refer to the references specified with each step to 
obtain additional information.
Table 2-1: PI-MAX4 System Installation Procedure (Sheet 1 of 2)
Action
For additional information, refer 
to…
1.
If the system components have not already been unpacked, unpack 
them and inspect their carton(s) and the system components for in 
transit damage.
, on 
2.
Verify that all system components have been received.
, on 
3.
If the camera will be used on a spectrograph, mount it to the 
equipment using the required adapter(s.)
, on 
4.
If the appropriate interface card is not already installed in the host 
computer, install it and its drivers.
, on 
5.
If the application software has not already been installed on the host 
computer, install it.
WinView User Manual, 
document number 4411-0046
WinSpec User Manual, 
document number 4411-0048
LightField User Manual, 
document number 4411-0125
6.
With the PI-MAX4 power supply and computer power turned OFF, 
connect the Ethernet cable (GigE) to the PI-MAX4 and the interface 
card in the host computer.
7.
Make the following connections based upon the type of system being 
setup:
Air-Cooled System
Plug the power supply into the rear of the camera and plug the 
power supply into the power source.
Liquid-Cooled System
Make the hose and power connections to the camera and plug the 
circulator into the power source. Add coolant if necessary. Turn 
on the circulator.