Princeton 4411-0139 ユーザーズマニュアル

ページ / 318
296
PI-MAX
®
 4 System Manual
Issue 4
4.
Position the 2.739” inner-diameter o-ring into the groove on the face of the camera. See 
.
Figure G-4: Positioning the 2.739” (ID) O-ring
5.
Using the four (4) screws removed in step 1, mount the spectroscopy-mount adapter to 
the front of the PI-MAX4.
6.
Position the smaller diameter o-ring into the groove on the face of the adapter. See 
.
Figure G-5: Positioning the 2.614” (ID) O-ring
NOTE:
A spare set of o-rings may be provided. When installing a new 
o-ring, be sure to apply a thin coat of vacuum grease before 
putting it in place. Lightly coating the o-rings with vacuum 
grease helps to insure an air-tight seal is formed between the 
camera and adapter, and between the adapter and the 
spectrograph sliding tube.
2.739” (ID) O-ring
In Groove
4411-0137_0177
2.614” (ID) O-ring
In Groove
4411-0137_0178