Emerson M21 ユーザーズマニュアル

ページ / 47
 
24 
MAN-0076 Rev 05 Millennium II 
December 07, 2012 
Net Safety Monitoring Inc 
 
4.2.1 Full calibration (Normal calibration) procedure 
Prior to attempting calibration read and understand the calibration procedure below. Also see 
or 
additional reference. 
 
The following calibration procedure should be followed to ensure an accurate correlation between the output signal 
and the gas concentration. For accurate performance, the Millennium II is calibrated using 50% span gas. The 
transmitter will however, allow some flexibility in the use of calibration gas with some sensors; calibration gas outside 
of 50 % span (10% - 60% span gas) will be allowed on specific sensor models (see specific sensor manual for details). 
The calibration gas value can be chosen by selecting it under “cal. gas value” in the main menu. A full calibration will 
take approximately 5 minutes to complete.   
 
 Ensure the transmitter is functioning properly as indicated by the status LED and current output.  
 
1.  Enter the main menu by selecting/activating any key to get the “enter main menu” prompt, then activate 
switch 1 to select “yes”. 
 
2.  When “Calibrate Sensor?” is displayed, activate the enter key (switch 3).  
 
3.   When “Calibrate Sensor #1?” is highlighted, activate the enter key (switch 3) if this is the sensor to be 
calibrated.  
 
4.  If sensor #2 is to be calibrated, select the down arrow key (switch 2) to scroll to “Calibrate Sensor #2?”  
 
5.  Select the desired sensor to be calibrated (1 or 2) by activating the enter key (switch 3).  
 
6.  Select “YES” with switch 1 to confirm the selection, and then apply clean air (zero gas) from canister when 
“Apply Clean Air”   is displayed. Ensure no contaminant gases are around if ambient air is being used. 
 
7.  Select “Z & Span” using switch 1 for normal (full) calibration. “Setting zero” will be displayed as the sensor 
is being zeroed.  
 
8.  Apply 50% calibration gas (* or % cal. gas value chosen) when prompted.  
 
9.  The display will show “Spanning” with the gas value (%LEL or PPM depending on the sensor) as the gas is     
detected.  
 
10.  Remove the calibration gas when “Remove Cal Gas” is displayed.  
 
11.  “Cal Complete” will be displayed when calibration is complete.  
 
12.  Apply zero gas (clean air) to purge system. This is particularly important when using long tubing. 
 
* Note:  Selectable calibration gas value (% cal. gas value) is only available for some sensor types. 
 
Warning
 
 
Always apply test gas after any calibration to verify accuracy; do a bump test after calibration. When 
applying test gas, make sure the system is bypassed to avoid unwanted shutdowns.