 Reference, Installation, and Operations Manual
Section 6: Meter operation
3-9000-743 Rev S
June 2013
Flow-condition pressure and temperature
 207
Data updates
When the Option Board is used, the Daniel Mark III Ultrasonic Gas Flow Meter samples the input 
analog signal(s) and updates the corresponding data point(s) 
(
LiveFlowPressureLiveFlowTemperature) once per second regardless of the input selection 
(disabled, live, or fixed).
Every five seconds, the meter updates the “in-use” flow-condition pressure and temperature 
values (
FlowPressureAbsFlowPressureFlowTemperature) depending upon the input 
selection, validity of the input data, and the selected data source upon alarm according to 
Table 6-6  Flow-condition pressure and temperature data source
Input Type
(EnablePressureInput or 
EnableTemperatureInput)
Data Validity
(
PressureValidity or 
TemperatureValidity)
Data Source Upon Alarm
(
FlowPOrTSrcUponAlarm)
“In-Use” Data Source
(
FlowPressure or 
FlowTemperature)
Disabled
N/A
N/A
“In-Use” value unchanged
Live
Valid
N/A
Average of live values
(
LiveFlowPressure or Live-
FlowTemperature)
Invalid*
Last good value
“In-Use” value unchanged
Fixed
Fixed data point
(SpecFlowPressure or Spec-
FlowTemperature)
Fixed
Valid
N/A
Fixed data point
(SpecFlowPressure or Spec-
FlowTemperature)
Invalid
Last good value
“In-Use” value unchanged
Fixed
Fixed data point
(SpecFlowPressure or Spec-
FlowTemperature)
*Live input can be invalid due to (a) one or more live values is/are at or outside the alarm limits, or (b) the input is being cali-
brated.