 Caution 
When the memory has been filled, the oldest block of data will 
be discarded to allow new data to be recorded. 
Measurement storage capacity can range between 28 hours and 2 years depending on the 
record period and enabled channels. Refer to Table 5-2, Approximate Data Storage 
Capacity. 
Table 5-2. Aproximate Data Storage Capacity 
Record Period 
Storage, 1 channel 
Storage, 2 channels 
1 second 
56 hours 
28 hours 
10 seconds 
20 days 
10 days 
1 minute 
18 weeks 
9 weeks 
5 minutes 
20 months 
10 months 
10 minutes 
24 months 
20 months 
30 minutes 
24 months 
24 months 
1 hour 
24 months 
24 months 
 
avu019.jpg 
Figure 5-18. Record Setting 
Use the A D buttons to select the setting to change. The B C buttons are used to change 
the setting parameters. Press the E button to save the new setting. Press the F button 
to cancel any changes and exit to the menu. 
Data View  
The DATA VIEW function allows viewing recorded data in NUMERIC or GRAPHIC 
format.