Fluke 5020A ユーザーズマニュアル

ページ / 142
5020A 
Users Manual 
5-26 
Sensor Alarm 
The SENSOR ALARM function allows the setting of the alarm parameters for each 
channel.  
To set alarm parameters, a channel number and measurement type (T or H) must first be 
selected using B C and ENTER buttons.  
 
avu036.jpg 
Figure 5-35. Sensor Alarm 
The PARAM parameter is selectable between 1/T, 1/H, 2/T, 2/H. The 1 and 2 refer to the 
channel selected, T is temperature and H is humidity. 
The A D buttons are used to select the menu option. The B C buttons are used to change 
the parameter setting. Press the E button to save the new setting. Press the F button 
to cancel any changes and exit to the menu. 
Once an option is selected, the parameters LOWER, L ENAB, UPPER, U ENAB, RATE, 
R ENAB, and SENSOR for the given channel (1 or 2) and measurement type (T or H) are 
displayed. The LOWER parameter is the lower limit alarm. The L ENAB parameter is 
used to turn the lower limit alarm ON or OFF. The UPPER parameter is the upper limit 
alarm. The U ENAB parameter is used to turn the upper limit alarm ON or OFF. The 
RATE parameter is the upper limit of the absolute value of the rate of change expressed 
in degrees/hour or %RH/hour. The R ENAB parameter is used to turn the rate alarm ON 
or OFF. The SENSOR parameter is used to turn the sensor fault alarm ON or OFF. 
The A D buttons are used to select the menu option. The B C buttons are used to change 
the parameter setting. Press the E button to save the new setting. Press the F button 
to cancel any changes and exit to the menu. 
System Alarm 
The SYSTEM ALARM function allows the battery or power alarms to be turned ON or 
OFF.  
When the BATTERY alarm is turned ON, an alarm occurs when the battery charge is 
low.  
When the POWER alarm is ON and a battery is installed, an alarm occurs when the 
power is disconnected. This alarm will continue to sound every 20 seconds until power is 
restored or the battery is discharged.