Digital Communications Interface
 
 
Commands
 
6
 
  
6-13 
  <hour> indicates a one or two digit number for the hour (0 to 23) is required. 
  <minute> indicates a one or two digit number for the minute is required. 
  <second> indicates a one or two digit number for the second is required. 
  <baud> indicates a valid baud rate setting is required. 
  For query commands, specifying the MIN, MAX, or DEF parameter causes the 
instrument to respond with the minimum, maximum, or default setting 
respectively. 
  For set commands, specifying the MIN, MAX, or DEF parameters causes the 
instrument to use the minimum, maximum, or default setting respectively. 
  Unrecognized commands or commands with incorrect syntax or invalid 
parameters generate error messages that are placed in the error queue (see 
SYST:ERR?). 
Alarm Commands 
ALARm:BATTery? 
Indicates whether a battery alarm event has occurred. 
Example: ALAR:BATT? 
Response: 1 
A value of 1 is returned if a battery alarm event has occurred. A value of 0 is returned if a 
battery alarm event has not occurred. 
ALARm:BATTery:ENABle? 
Returns the state of the battery alarm enable. 
Example: ALAR:BATT:ENAB? 
Response: 0 
A value of 1 is returned if the battery alarm is enabled. A value of 0 is returned if the 
battery alarm is disabled. 
ALARm:BATTery:ENABle <bool> 
Sets the state of the battery alarm enable. 
Example: ALAR:BATT:ENAB 1 
The <bool> parameter turns the battery alarm enable on (1 or ON) or off (0 or OFF). This 
command may be password protected (see Password Commands). 
ALARm:BEEP? 
Returns the state of the audible alarm enable. 
Example: ALAR:BEEP? 
Response: 0 
A value of 1 is returned if the alarm beep is enabled. A value of 0 is returned if the alarm 
beep is disabled.