PI-MTE System Manual 
Version 4.C 
4.  After the vacuum is fully vented, clamp or 
support the camera with the CCD array 
looking up. 
5.  Loosen, by one turn, each of the eight (8) 
socket head screws on the face of the nose. 
6.  After all of the 8 screws have been loosened 
one turn, remove all 8 screws. 
7.  "Rock" the nose slightly to be sure the 
elastomer gasket has released. 
8.  Be careful in performing the next step. Unless 
you are careful, the CCD array and/or the bond 
wires could be damaged.  Such damage is not 
covered under warranty. 
9.  Remove the nose in a direction perpendicular 
to the CCD plane. Do not slip the nose 
sideways. 
10.  Remove the elastomer gasket. 
11.  Store the nose and gasket in a secure location 
for future use. 
12.  Being very careful, mount the open nose with the 
8 screws previously removed. (The elastomer 
gasket is not required.) In a criss-cross pattern, 
tighten the screws to 15 in-lbs. 
13.  Cover the nose opening whenever the camera 
is not in the vacuum chamber.