TM 11-6625-1614-15
Model 410C
SECTION II
INSTALLATION
.
2-1. INSPECTION.
2-2. 
‘This 
instrument was carefully inspected both
mechanically and electrically, before shipment. It
should be physically free of mars or scratches and in
perfect electrical order upon receipt. To confirm
this, the instrument should be inspected for physical
damage in transit. Also, check for supplied acces-
sories, and test the electrical performance of the in-
strument using the procedure outlined in Paragraph
5-5 Performance Checks. If there is any
damage or deficiency, refer to
paragraph 1-A.3.
2-3. INSTALLATION.
2-4. The     Model 410C is transistorized except for
one vacuum tube and requires no special cooling.
However, the instrument should not be operated where
the ambient temperature exceeds 55° C (140° F).
2-5. RACK MOUNTING.
2-6. The Model 410C is a submodular unit designed
for bench use. However, when used in combination
with other submodular units, it can be bench and/or
rack mounted.
The    Combining Cases and Adapter
Frame are designed specifically for this purpose.
2-7. MODELS 1051A AND 1052A COMBINING CASES.
The Combining Cases are full-module unita which
accept various combinations of submodular
units.
Beinga full width unit, it can either be bench or rack
mounted. An illustration of the Combining Case is
shown in Figure 2-1. Instructions for installing the
Model 410C are shown in Figure 2-2.
2-8. RACK ADAPTER 
FRAME 
(      Part No. 5060-
0797). The adapter frame is a rack mounting frame
that accepts various combinations of submodular
units. It can be rack mounted only.
An illustration
a. Place the adapter frame on edge of bench as
b. Stack the submodular units in the frame as
c. Place spacer clamps on the two end instruments
the frame.
d. Insert screws on either side of frame, and
tighten until submodular instruments are tight in the
frame.
e.
The compiete assembly is ready for rack
mounting.
01556-2
2-9. THREE-CONDUCTOR POWER CABLE.
2-10. 
To protect operating personnel, the National
Electrical Manufacturers’ Association (NEMA) re-
commends that the instrument panel and cabinet be
grounded.
All Hewlett-Packard instruments
are
equipped with a three-conductor power cable
which
grounds the instrument 
when plugged into an appropri-
ate receptacle.
2-11. To preserve the protection feature when oper-
ating 
ihe instrument from a two-contact outlet, use
three - prong to two - prong adapter and connect the
green pigtail on the adapter to ground.
2-12. PRIMARY POWER REQUIREMENTS.
2-13. 
The Modei 410C can be operated from either
115 or 230 volts, 50 to 1000 cps. The instrument can
be easily converted from i 15- to 230- volt operation.
The LINE VOLTAGE 
switch, S4 a two-position 
slide
switch located 
at 
the rear of the instrument, selects
the mode of AC operation. The line 
voltage from
which the instrument is set to operate appears on the
siider of the switch. A 0.25-ampere, slo-blo fuse is
used for both 115- and 230-volt operation.
Figure 2-3. Adapter Frame Instrument Combination