Command Reference 9-101
:STATus :OPERation :CONDition?
80B/81/85
Read Operation Status Condition Register
Reads out the contents of the operation status condition register. This register re-
flects the state of the measurement process. See figure below. Note that bits 0 to
3, 7, and 9 to 15 are not used.
Returned Format:
<Decimal data> = the sum (between 0 and 368) of all bits that are true. See table below:
Bit No.
Weight
Condition
8
256
Not Measurement
6
64
Waiting for bus arming
5
32
Waiting for triggering and/ or external arming
4
16
Measurement
Complies to standards:
SCPI 1991.0, confirmed