Programming Site Data
4-45
April  2004
Entering the Digital Output (cont.)
If you entered Test Points:
The prompt asks for the signal type from two choices: transmit or 
receive.
a. Press 
[ENTER]
 to open the drop-down menu.
b. Scroll to the required signal.
c. Press 
[ENTER]
 to confirm your selection.
If you entered Gate Input:
1. The first prompt asks for the gate active.
a. Press 
[ENTER]
 to open the drop-down menu.
b. Scroll to the gate, either Contact Open or Contact Closed.
c. Press 
[ENTER]
 to confirm your selection.
2. The remaining prompt asks for the mode, automatic or 
manual.
a. Press 
[ENTER]
 to open the drop-down menu.
b. Scroll to the required mode.
c. Press 
[ENTER]
 to confirm your selection.
Stopwatch Totalizer
Through the Gate Input option, you can implement the  
Stopwatch Totalizer function to measure totals manually. To set 
up the Totalizer function:
1. At the Gate Active prompt, select Contact Closed.
2. At the Mode prompt, select Manual, and press 
[ENTER]
.
The stopwatch icon (see page 2-11) appears in the system tray. To 
start or stop the function, press the minus (-) key on the keypad.
You have completed entering parameters in the Digital Output 
option.
To confirm the entries and return to Operate mode, press 
[F3]
 
(OK). 