GE PT878GC ユーザーズマニュアル

ページ / 308
Servicing the PT878GC
9-13
April 2004
Calibrating the Thickness Gauge Transducer (cont.)
a. Press 
[ENTER]
 to open the text box.
b. Use the numeric keys to enter the length.
c. Press 
[ENTER]
 to confirm the entry.
4. Press 
[ENTER]
 (Calibrate button) to prepare for calibration. 
Then apply couplant B to the block surface, and hold the 
transducer against the block. 
5. When you have the transducer firmly pressed against the 
block, press [ENTER] (Ready button) again to perform the 
calibration.
6. Hold the transducer steady until the Set button appears. Press 
[ENTER]
 to store the calibration. 
IMPORTANT: Be sure to hold the transducer steady until the 
GE Panametrics wait cursor disappears.
7. If you have chosen dual calibration, repeat Steps 3 through 6 
above for Block 2.
8. The PT878GC asks for confirmation of the calculated and 
current values. 
Press the [
S
] arrow key to move to the Retry box and press 
[ENTER]
 to recalculate the value, or
Press 
[ENTER]
 on “Set” to commit the calculated value.
Once you have completed calibration, the transducer should 
remain calibrated until the meter memory is cleared or a dramatic 
shift in ambient conditions occurs.
To leave the Zero window:
Press the [
W
 ] or [
X
] arrow key to move to another tab to 
take readings or perform a velocity calibration.
Press 
[F2]
 (Cancel) to return to Operate Mode without 
confirming the sound speed value.
Press 
[F3]
 (OK) to confirm the new value. The PT878GC 
returns to Operate Mode.