Operation 
Teledyne API M751H High Performance Zero Air Generator 
 
26 
 
07773A DCN6833 
 
 
Provision for voltage configuration is via a jumper plug (J5). All electrical connections 
are made by quick release connectors to aid in servicing.  Refer to Figure 4-4 for the 
layout and to Appendix B for the interconnect and other drawings. 
The 751H Control PCB is microcontroller based and provides the following functions: 
  Cycles the four-way valve of the Regenerative Dryer. 
  Cycles the water drain solenoid valve. 
  Cycles the relief valve in response to the pressure sensor.  
  Cycles the Regenerative Dryer four-way valve and momentarily opens the water 
drain solenoid valve at compressor startup. This reduces the pressure at the 
compressor outlet to facilitate compressor starting. 
LED  D13  flashes  at  1  second  intervals  as  an  indication  that  the  controller  is 
functioning, and a watchdog timer is enabled to prevent any power line disturbances 
from halting the processor. 
 
Figure 4-4:  Control Board Layout