Omega OMB-DAQ-2416_4AO ユーザーズマニュアル

ページ / 53
 
25 
Chapter 5 
Specifications 
All specifications are subject to change without notice. 
Typical for 25 °C unless otherwise specified. 
All specifications apply to all temperature and voltage input channels unless otherwise specified. 
Specifications in italic
 text are guaranteed by design. 
Analog input 
Table 1. General analog input specifications 
Parameter 
Conditions 
Specification 
A/D converter type 
 
ADS1256, 24-bit Sigma Delta 
A/D data rates 
 
3750 samples per second (S/s), 2000 S/s, 1000 S/s, 500 S/s, 100 S/s, 
60 S/s, 50 S/s, 25 S/s, 10 S/s, 5 S/s, 2.5 S/s 
Throughput 
 
  Single channel: 2.5 Hz to 1102.94 Hz, software selectable 
  Multiple channels: 0.16 Hz to 1102.94 Hz, software selectable 
See Table 18 and Table 19 for details. 
Number of channels 
 
Up to 32 channels individually software-configurable as single-
ended or differential  
Thermocouples require differential mode. For each channel 
configured as differential, you essentially lose one single-ended 
channel. You can add channels by connecting to an 
OMB-AI-EXP32. 
Input isolation 
 
500 VDC minimum between field wiring and USB interface 
Channel configurations 
 
Temperature sensor input, s/w programmable to match sensor type 
Voltage input 
Input voltage range 
Thermocouple mode 
±0.078125 V 
Voltage mode (Note 1) 
±20 V, ±10 V, ±5 V, ±2.5 V, ±1.25 V, ±0.625 V, ±0.3125 V, 
±0.15625 V, ±0.078125 V, software-configurable 
Absolute maximum 
input voltage 
CxH-CxL relative to 
GND  
  ±30 V maximum (power on) 
  ±10 V maximum (power off) 
Input impedance 
 
  2 GΩ (power on) 
  390Ω (power off) 
Input leakage current 
 
±10.6 nA 
  Input voltage >±30 V 
(power on/off) 
±1 µA maximum 
Input capacitance 
 
590 pf 
Maximum working 
voltage (signal + 
common mode) 
Voltage mode: ±20 V 
range 
±20.01 V maximum 
Voltage mode: all other 
voltage input ranges 
±10.25 V maximum 
Common mode 
rejection ratio (Note 1) 
Thermocouple mode, 
(f
IN 
= 60 Hz) 
110 dB 
Voltage mode,(f
IN
 = 
60 Hz, all input ranges) 
90 dB 
ADC resolution  
 
24 bits 
Crosstalk 
Adjacent channels 
100 dB 
Input coupling  
 
DC 
Channel gain queue 
Up to 64 elements 
Software-configurable channel and range 
Warm-up time 
 
45 minutes minimum 
Open thermocouple 
detect 
 
Automatically enabled when the channel is configured for a 
thermocouple sensor.