Omega OMB-DAQ-2416_4AO ユーザーズマニュアル

ページ / 53
OMB-DAQ-2416-4AO User's Guide 
Specifications 
40  
Analog voltage output 
Unused VDACx output channels should be left disconnected.  
The OMB-DAQ-2416-4AO output voltage level defaults to 0 V whenever the host PC is reset, shut down or 
suspended, or if a reset command is issued to the device. 
The duration of the output transient depends highly on the enumeration process of the host PC. Typically, the 
output of the OMB-DAQ-2416-4AO is stable after two seconds.  
Table 20. Analog voltage output specifications 
Parameter 
Conditions  
Specifications 
Digital-to-analog 
converter 
 
DAC8555 
Number of channels 
 
Resolution 
 
16 bits 
Output ranges 
Calibrated 
±10 V 
Uncalibrated 
±10.05 V, software configurable 
Output transient 
Host PC is reset, powered on, suspended or a 
reset command is issued to device (Note 6) 
Duration: 2 s 
Amplitude: 2 V p-p 
Initial power on 
Duration: 50 ms 
Amplitude: 5 V peak 
Differential 
non-linearity 
 
±0.25 LSB typical 
±1 LSB maximum
 
Output current 
VDACx pins  
±3.5 mA maximum 
Output short-circuit 
protection 
VDACx connected to AGND 
Unlimited duration  
Output coupling 
 
DC 
VDACx readback 
 
Each VDACx output can be independently 
measured by the onboard A/D converter. 
Software selectable 
Power on and reset 
state 
 
DACs cleared to zero-scale: 0 V, ±50 mV 
Output noise 
 
30 µVrms 
Settling time 
To rated accuracy, 10 V step 
45 µs 
Slew rate 
 
1.0 V/µs 
Throughput 
Single-channel 
1000 S/s maximum, system-dependent 
Multi-channel 
1000 S/s #ch maximum, system-dependent 
Table 21. Calibrated absolute accuracy specifications 
Range 
Accuracy (±LSB) 
±10 V 
16.0 
Table 22. Calibrated absolute accuracy components specifications 
Range 
% of reading 
Offset (±mV) 
Temp drift (%/°C) 
Absolute accuracy at FS (±mV) 
±10 V 
±0.0183 
1.831 
0.00055 
3.661 
Table 23. Relative accuracy specifications 
Range 
Relative accuracy (±LSB) 
±10 V 
4.0 typical