3M DL ユーザーズマニュアル

ページ / 168
29    Basic setups 
 
  Measurement setup (for community noise) 
 
 
Models SE and DL 
 
• 
CNEL  Press Enter       to select data field. Press Enter       
repeatedly to toggle between On/Off setting. 
• 
L
c-a
 – Press Enter       to select data field. If marked “N/A,” see “C-
A measurement settings.” 
• 
TAKTMX ~ Press Enter        one or more times to select 3 sec., 5 
sec., or Off. 
6.  To change other measurement settings, repeat steps 4-5.  
 
7.  When finished, press On/Off/Esc       to exit the screen.  
  
The following table explains your selections in the Measures screen 
(Please see Appendix B, Glossary of Terms for more details.) 
Measurements 
Explanation  
 
L1-L4 
(Exceedance) 
L1-L4 are types of exceedance levels.  This is defined as the 
level exceeded by the measured noise level for an identified 
fraction of time.  Exceedance levels may be calculated for 
many time fractions (or percentages).  
 
•  Optional Settings:  Set to “Off” or to a percentage value of 
your choice (1-99%).   
LDN  
Representing the Day/Night sound level, this measurement is 
24-hour average sound level where 10 dB is added to all of the 
readings that occur between 10pm to 7am.  (Please see 
glossary of terms for more details).   
 
•  Optional Settings:  Set to “Off” or “On”. 
CNEL 
Community Noise Exposure Level.  The accumulated exposure 
to sound measured in 24-hour sampling interval and artificially 
boosted during certain hours.   
 
Optional Settings:  Set to “Off” or “On”. 
L
c-a
 
This measurement (also commonly called C-A measurement) is 
only available under certain conditions.  Information about 
viewing and changing these meter settings is discussed under 
the next section, “C-A measurement settings”.  Optional 
Settings:  Set to “On” and “Off”.  See C-A section below for 
parameter settings. 
TAKTMX 
A time average over a sampling that uses the highest  level 
occurring during the preceding Taktmaximal interval (either 
three or five seconds), and assumes it was present for the 
whole Taktmx interval.  
 
Optional Settings:  Set to “Off”, “3sec”, or “5sec” 
 
 
Table 3-2:  Measures screen explained