Tektronix VX4101A ユーザーズマニュアル

ページ / 545
SCPI Commands for the VX4101A
VX4101A MultiPaq Instrument User Manual
3–27
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:FIRe <command trigger> 
Generates one of five software triggers: COMMAND0 through COMMAND4.
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:TIMer <timer seconds> 
Sets or queries the value of the period trigger source. Note that this is a global
class command and there is only one periodic timer for the VX4101A. Any time
you program the TRIG:TIM command, the period of the timer trigger will be
changed for all instruments using it as a source.
Table 3–2: Trigger Resolution (in 
s)
 
Lower Bound
Upper Bound
Resolution
2
< 510
2
510
< 1275
5
1275
< 2040
8
2040
< 6375
25
6375
< 8160
32
8160
< 12750
50
12750
< 25500
100
TRIGger ([:SEQuence1]|:STARt)[LAYer]:TIMer 
This query returns the value of the periodic trigger source in seconds.
TRIGger ([:SEQuence1]|:STARt)[:LAYer]:FIRe <0-4> 
Command
Response
TRIG:SOUR:CAT?
HOLD, IMMEDIATE, BUS, TTLTRG0, TTLTRG1, TTLTRG2,
TTLTRG3, COMMAND0, COMMAND1, COMMAND2,
COMMAND3, COMMAND4, TIMER, SUREPATH, DMM,
COUNTER, CTR_EXTARM, DAC, DIGI, DIGO, HANDSHAKE
INST:SEL DMM
CONF:VOLT:DC
TRIG:SOUR COMM0
INIT
FETC:COUN?
0
TRIG:FIR0
FETC:COUN?
1
FETC?
–6.03720E+00
Command Description
Query Response
Examples