Fluke 9011 ユーザーズマニュアル

ページ / 52
38530
Current ALPHA setting
To change the ALPHA setting, press “UP” or “DOWN”.
38600
New ALPHA setting
To accept the new setting, press “SET”. Press “EXIT” to continue without stor-
ing the new value.
S
Accept the new ALPHA setting
6.14.4
DELTA
This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for
best instrument accuracy.
deLta
Flashes for one second and then the DELTA setting is
displayed
1.5070
Current DELTA setting
To change the DELTA setting, press “UP” or “DOWN”.
1.6320
New DELTA setting
To accept the new setting, press “SET”. Press “EXIT” to continue without stor-
ing the new value.
S
Accept the new DELTA setting
6.14.5
BETA (cold side only)
This probe parameter relates to the higher order nonlinearity of the sensor be-
low 0°C. The value is set at the factory for best instrument accuracy.
bEtA
Flashes for one second and then the BETA setting is
displayed
0.111
Current BETA setting
To change the BETA setting, press “UP” or “DOWN”.
0.100
New BETA setting
To accept the new setting, press “SET”. Press “EXIT” to continue without stor-
ing the new value.
9011 Dual-Well Calibrator
User’s Guide
34