SECTION 4—COMPRESSOR OPERATION
29
Turning the Compressor On
NOTE: For this procedure, refer to FIGURE 4.5 on page 30.
1. Make sure the concentrator is On. Refer to the 
concentrator Owner’s Manual.
CAUTION
The concentrator flow rate to the patient MUST be set at 
maximum: 3 L/min. or less for Platinum 5, XL and 5 
L/min. for Platinum 10. Otherwise, the oxygen level to the 
compressor will be <90% and the compressor will not fill 
the cylinder.
2. Set the concentrator flow rate to the patient to 3 L/min. or 
less for Platinum 5, XL and 5 L/min. for Platinum 10. Refer 
to the concentrator Owner’s Manual.
3. Make sure the cylinder is connected to the compressor. 
4. Push the power switch on the control panel to the On ( ) 
position.
5. Examine the control panel. The following sequence of 
events should occur:
A. 0-3 minutes - The O
2
 Below Normal (YELLOW) light 
will be lit while the compressor warms up.
B. After three minutes - The FILLING (GREEN) light will 
be on while the cylinder is filling.
C. The FULL (GREEN) light will be on when the cylinder 
is finished filling. Proceed to STEP 6.
NOTE: The O
2
 BELOW NORMAL (YELLOW) light may come on 
if the oxygen level from the concentrator has not yet reached or drops 
below 90%. The compressor will begin or resume filling when the