SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES
34
SECTION 6—TROUBLESHOOTING/ 
CYLINDER FILL TIMES
TROUBLESHOOTING
PROBLEM
SOLUTION
No indicator lights on 
when compressor turned 
on.
1. Turn compressor Off.
2. Check that power cord is properly plugged 
into compressor and wall outlet
.
3. Turn the compressor On.
5. If performing STEPS 1-4 does not cause any 
indicator lights to come on, contact your home 
health care provider or Invacare for service.
RED Light ON
1.Check the cylinder gauge to see if it is FULL. 
Cylinder full - Remove the cylinder and turn the 
unit Off. Perform the Cylinder Prefill Inspection 
on page 16 with a cylinder that is NOT FULL.
Cylinder not full - Proceed to STEP 2. 
2. Check that the cylinder is connected properly. 
Turn the compressor Off. Remove the cylinder 
and reinstall. Turn the compressor On.
3. If cylinder is not full and is connected properly, 
the compressor may have internal failure. 
Contact your home health care provider or 
Invacare.
Compressor cycles ON 
and Off
or
O
2
 BELOW NORMAL
(YELLOW) and FILLING 
(GREEN)
lights are alternating 
or
O
2
 BELOW NORMAL 
(YELLOW) 
light goes On and Off
1. Check the concentrator liter flow setting is at 3 
L/min. or less.
If necessary, adjust the liter flow setting (refer to 
the concentrator owner’s manual). Wait at least 
20 minutes for the GREEN light to appear. If it 
does not appear, proceed to STEP 2.
2. The concentrator may need service, contact 
your home health care provider or Invacare.