Teledyne API Ultrafine Particle Monitor - Model 651
Maintenance, Service, and Troubleshooting
07506C DCN6727
81
C h a n g i n g t h e F i l t e r s
The Model 651 contains three filters. The filters should be replaced
as part of the annual service.
as part of the annual service.
To replace a filter, follow these instructions:
1. Turn off the power to the Model 651.
2. Remove the instrument cover.
3. Remove the filter from the filter clip.
Transport
Flow Filter
Flow Filter
Auxiliary
Flow Filter
Flow Filter
Sample Flow
Filter
Filter
Figure 9-5
Location of Filters
Location of Filters