Fluke A20M-AF ユーザーズマニュアル

ページ / 148
 
4. REMOTE OPERATION 
 
Page 101 
© 2007 DH Instruments, a Fluke Company 
4.6 
IEEE STD. 488.2 COMMON AND STATUS PROGRAM 
MESSAGES 
The RPM4 supports a set of commands that are common to all instruments conforming to IEEE Std. 488.2.  
These command make it easy to perform basic function for any device that supports these commands.   
These commands also cover the status reporting commands.  See Section 4.5 for details on the status 
registers mentioned in these commands. 
Table 19.  Program message list 
CLS 
Clear all of the status & event structures. 
ESE 
Read or set the Standard Event Status Enable Register. 
ESR 
Read the Standard Event Status Register. 
IDN 
Identify the RPM4 version, range, and serial number. 
OPC 
Set the operation complete bit when all operations have completed. 
OPT 
Read the list of installed RPM4 options. 
RST 
Reset the RPM4  settings to factory settings. 
TST 
Read the power on self test status. 
SRE 
Read or set the Service Request Enable Register. 
STB 
Read the Status Byte Register. 
RSE 
Read or set the Ready Status Enable Register. 
RSR 
Read the Ready Status Register. 
 
4.6.1 
PROGRAM MESSAGE DESCRIPTIONS 
CLS 
Purpose 
Clear all of the status & event structures. 
Command 
∗CLS” 
Remarks 
This program message clears the following events and status registers: 
Standard Byte Register (STB) 
Standard Event Status Register (ESR) 
Error Queue 
Pending OPC operations 
Example 
 
Sent: 
Reply: 
∗CLS” 
none 
 
ESE 
Purpose 
Read or set the Standard Event Status Enable Register. 
Command 
Query 
∗ESE n” 
∗ESE?” 
Default 
∗ESE 0” 
Arguments 
n: 
’0 to 255’  This is the decimal representation of the bit(s) to enable. To enable 
the PON and QYE bits, the argument would be 128 + 4 = 132. 
Remarks 
The Standard Event Status Enable register determines which bits in the standard Event 
Status Register are enabled and included in the Status Byte Register (ESB bit), and can 
assert the SRQ line.  The reply is in decimal numeric form. 
Example 
(enhanced) 
Sent: 
Query reply: 
∗ESE=128”(enables the PON bit)  
“128” (no reply if IEEE-488) 
Errors ERR# 
6: 
n is not valid. 
 
ESR 
Purpose 
Read the Standard Event Register. 
Command 
∗ESR?” 
Remarks 
The Standard Event Register contents are cleared after reading.  The reply is in decimal 
numeric form.  
Example 
(enhanced) 
Sent: 
Reply: 
∗ESR?”  
“20” (the QYE and EXE bits are set)