RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 106 
If the RPM4 is used only in gauge mode, Autozeroing in absolute mode is not 
necessary and vice-versa.  Gauge and absolute mode Autozeroing are 
completely independent one from the other. 
See Section 3.5.1 for complete information on Autozeroing of RPTs. 
5.2.1.2 
ABSOLUTE MODE AUTOZERO 
When operated in absolute mode, a Q-RPT should be Autozeroed about once a 
month or when the RPM4 has been exposed to a temperature change exceeding 
15 ºC (36 ºF).  Whether or not AutoZeroing is necessary can also be evaluated by 
comparing the reading of the RPM4 Q-RPT to the reading of a barometer when 
the RPM4 is vented at atmospheric pressure.  The barometer should have 
measurement uncertainty of 0.04 psi (0.28 kPa) or better.  For the Lo Q-RPT, if 
the disagreement is greater than 0.05 psi (0.35 kPa), the Q-RPT should be 
rezeroed against the barometer.  For the Hi Q-RPT, if the disagreement is 
greater than 0.15 psi (1 kPa), it should be rezeroed against the Lo Q-RPT or 
against the barometer. 
Follow the procedures described in Section 3.3.9.2 to perform absolute mode 
Autozeroing. 
5.2.2 
ADJUSTMENT OF THE ON-BOARD BAROMETER 
 PURPOSE 
To adjust the output of the on-board barometer (see Section 3.5.7.1). 
 PRINCIPLE 
The on-board barometer output can be adjusted using PA and PM values following the same 
principles as for the Q-RPTs (see Section 5.3.1.1).  
Since the on-board barometer is not a source of traceable pressure values, it does not have 
to be formally calibrated.  It should be offset to agree with a reference barometer.  To offset 
the barometer, change the barometer pressure adder (PA). 
 
  The on-board barometer is used only for measuring changes in atmospheric pressure over 
short periods of time (see Section 3.2.3.  RPM4 measurement accuracy does NOT depend on the 
absolute accuracy of the on-board barometer. 
 
 OPERATION 
To view or edit the values of PA and PM for RPM4’s on-board barometer, press [SPECIAL]
<8cal>,  <3barometer>.  Pressing [ENT] steps through displays of the calibration date 
[YYYYMMDD] and PA and PM.  In edit mode, the values can be edited.  Pressing [ENT] 
after the last screen activates the edited values. 
 
   To view the current output of the on-board barometer, press [SPECIAL], <7Internal>, 
<3baro> from the main run screen. 
 
  A pressure standard may be connected to the on-board barometer by connecting to one of 
the ATM ports on the RPM4 rear panel Q-RPT modules (10-32 UNF).  The operating span of the 
barometer is 10 psi (70 kPa) to 16 psi (110 kPa).